ECG/RESP Monitoring
ST ANALYSIS Pick this item to access the ST ANALYSIS menu, the detailed information
about the menu is to be discussed in the following section.
ARR ANALYSIS Pick this item to access the ARR ANALYSIS menu, the detailed
information about the menu is to be discussed in the following section.
OTHER SETUP
Pick this item to access the ECG SETUP menu as shown below:
MEDECG 5ETUF
ECG CAL
ADJUST UAUE FOS »
BEAT UOL DEFAULT »
PACE OFF
NOTCH OFF
Select the sound volume of the
heart beat.
EXIT
Figure 11-5 ECG SETUP menu
In this menu, following functions are available:
BEAT VOL
Four selections are available: OFF, LOW, MED, HIGH. HIGH indicates maximum volume.
OFF indicates no sound.
PA CE
"ON" detected signal will be marked by a " I " above the ECG waveform.
"OFF" mean the pacemaker analysis function is disabled.
14 Note 14
Note
If monitoring a patient with pacemaker, set "PACE" to ON, otherwise, set it to OFF.
If "PACE" is ON, the system will not perform some types of ARR analysis and ST
analysis. For detailed information, refer to the section: ARR ALARM.
• NOTCH:
The control switch for 50/60HZ notch waveform of ECG module. Three selections are
available, which are 50HZ, 60HZ and OFF. If you select 50HZ or 60HZ, the system can
respectively filter 50HZ or 60HZ industrial frequency interference.
• ECG CAL
Pick this item to start ECG calibrating process. Picking this item again or change lead
name on the screen can end calibrating process.
11-92