Mindray MEC-1000 Hastabaşı Monitörü
NIBP Monitoring CUFF TYPE ERR Cuff type does not LOW Select appropriate cuff type comply with the patient type. NIBP TIME OUT Measuring time has HIGH Measure again or use other exceeded 120 seconds measuring method. (adult) or 90 seconds (neonatal). Check the cuff. Make sure that the MEASURE FAIL patient under monitoring is HIGH motionless. Measure again. Prompt message: (display in the prompt area below NIBP value) Message Cause Alarm Level Manual measure During manual measuring mode. No alarm Cont measuring During continuous measuring mode. Auto measuring During automatic measuring mode. Please start After selecting interval between measurements in MENU Meas over Press START/STOP key during measuring to stop measurement Calibrating During calibrating Cal over Calibration over Pneu testing During pneumatic test Pneu test over pneumatic test over Resetting Resetting process after NIBP module is loaded Resseting In NIBP resetting process triggered that you start manufally Reset failed NIBP module reset failed 13.5 Maintenance and Cleaning Warning A Do not squeeze the rubber tube on the cuff. Do not allow liquid to enter the connector socket at the front of the monitor. Do not wipe the inner part of the connector socket when cleaning the monitor. When the reusable cuff is not connected with the monitor, or being cleaned, always place the cover on the rubber tube to avoid liquid permeation. 13-12