ECG/RESP Monitoring
Figure 11-16 RESP SETUP
RESP alarm setting
ALM: pick "ON", the system will give alarm prompt and save alarm information when RESP
alarm occurs. Pick "OFF", the system will not give alarm and instead display a & beside
"RESP".
ALM REC: pick "ON", the system will command the recorder to output the alarm
information when RESP alarm occurs.
ALM LEV: selectable from HIGH, MED and LOW. Level HIGH represents the most serious
case.
ALM HI: used to set up the upper alarm limit.
ALM LO: used to set up the lower alarm limit.
The system will give RESP alarm once the RESP value goes above the upper alarm limit or
below the lower alarm limit. RESP alarm limits:
Max. RR HI Min. RR LO Step
RESP ADU 120 0 1
RESP NEO/PED 150 0 1
APNEA ALM: used to set the standard of judging an apnea case. It ranges from 10 to 40
seconds with each turn of knob being 5s.
SWEEP: Available options are 6.25, 12.5 and 25.0 mm/s.
WAVE AMP: used to set up the magnified display of RESP waveform. The selections are
0.25, 0.5, 1, 2, 3, 4, 5.
HOLD TYPE: options are AUTO and MANUAL. When it is AUTO mode, HOLD HI and
HOLD LO menus cannot be used and the monitor automatically calculates the RESP
RATE. When it is MANUAL, you can use HOLD HI and HOLD LO items to respectively
adjust the two dotted lines in the RESP waveform area. The positions of these two dotted
lines are used to calculate the upper and lower RESP alarm limits.
HOLD HI and HOLD LO: When the HOLD TYPE is MANUAL, you can use the knob to
pick either HOLD HI or HOLD LO and turn the knob to respectively adjust the two dotted
lines in the RESP WAVEFORM area. The positions of these two dotted lines are used to
calculate the upper and lower RESP limits.
11-107