Mindray MEC-1000 Hastabaşı Monitörü
ECG/RESP Monitoring TACHY All patients 5 consecutive QRS TACHY User-sele complex, RR interval is ctable less than 500ms. BRADY All patients 5 consecutive QRS BRADY User-sele complex, RR interval is ctable longer than 1.5s. BEAT MISS Without When HR is less than 100 BEAT MISS User-sele pacemaker beats/min., no heart beat is ctable tested during the period 1.75 times of the average RR interval; or When HR is larger than 100 beats/min., no beat is tested with 1 second. PNP With No QRS complex and PNP User-sele pacemaker pacing pulse appear during ctable the period 1.75 times of the average R-R interval (only considering patients with pacemaker). PNC With When pacing pulse is PNC User-sele pacemaker available, no QRS exists ctable during the period 1.75 times of the average RR interval (only considering patients with pacemaker.) Patient type: All patients: refers to perform Arr. analysis on patients either with pacemakers or without pacemakers. Without pacemaker: refers to perform Arr. Analysis only on the patients without pacemakers. With pacemaker: refers to perform Arr. Analysis only on the patients with pacemakers. Prompt message: Message Cause Alarm Level ARR LEARNING The QRS template building required for Arr. Analysis is in No process. alarm A Note A The system displays the Arrhythmia name in the Alarm Message area. 11-104