NIBP Monitoring
■ NIBP alarm setups
ALM: pick "ON", the system will give alarm prompt and store alarm information once
NIBP alarm occurs; pick "OFF", the system will not give any alarm information and
instead display a & beside "NIBP".
ALM LEV: selectable from HIGH, MED to LOW. HIGH represents the most serious
case.
ALM REC: pick "ON", the system will command the recorder to output the alarm
information once NIBP alarm occurs.
SYS ALM HI, SYS ALM LO, MEAN ALM HI, MEAN ALM LO, DIA ALM HI, DIA ALM LO
are for you to set up the alarm limit for each type of pressure. NIBP alarm is activated
when the pressure exceeds the upper alarm limit or falls below lower alarm limit.
NIBP alarm limits:
Adult Mode
SYS 40-270 mmHg
DIA 10-215 mmHg
Mean 20-235 mmHg
Pediatric Mode
SYS 40-200 mmHg
DIA 10-150 mmHg
Mean 20-165 mmHg
Neonatal Mode
SYS 40-135 mmHg
DIA 10-100 mmHg
Mean 20-110 mmHg
■ UNIT
Options are mmHg and kPa.
■ INTERVAL
Time interval for AUTO measurement. Available selections are:
1/2/3/4/5/10/15/30/60/90/120/180/240/480 minutes. After selecting the INTERVAL, the
system will display the prompt "Please press START button" in the NIBP prompt area. At
this time, you can press the NIBP button on the front panel to start the first AUTO
measurement. Pick the [MANUAL] item in the list of INTERVAL and the system will end
AUTO measurement and restore MANUAL measurement mode.
■ INFLATION
Pick this item to choose the initial pressure when inflate the cuff next time. There are
different selections of the initial pressure.
Default Default initial Options for Initial value in NIBP MANUAL
value menu
(mmHg/kPa) (mmHg/kPa)
FACTORY DEFAULT 150 80/100/120/140/150/160/180/200/220/240
ADU CONFIG
FACTORY DEFAULT 100 80/100/120/140/150/160/180/200
PED CONFIG
13-127