What to do if
problem
Heart rate
symbol
Low battery
indicator
Measuring
error
Error
Reason
Appears in the measurement
condition and flashes when
pulse is detected.
Appears when the battery
voltage is excessively IoW or
the positions of the
batteries are incorrect.
Appears When the
accurate blood pressure and
pulse could not be
obtained.
Solution
• Measurement in progress, remain quiet
• Replace all four batteries With new ones. Insert
the batteries in the correct positions, Be aware
Of +/- positions.
• Press "Start/Stop" button again and remeasure.
• Check if cuff is wrapped according to
instructions.
• Check that there are no kinks in tube.
• Check palm if exerting effort.
• Check if talking or moving during
measurement.
• Check if posture is correct.
0297
IMPORTANT
Read the operating instructions.
Classification:
• Internally powered equipment
• Type BF equipment
• IPXO
• Not suitable for use in the presence of flammable anæsthetic mixture with air,
oxygen or nitrous oxide
• Continuous operation with short-time loading
Operating temperature
Storage temperature
This product conforms to the provisions of the EC directive 93/42/EEC (Medical
Device Directive). This device conforms to the following standards:
• EN 60601-1: 1990 + Al: 1993 + A2: 1995 A13: 1996 General requirements for
safety.
• EN 60601-1-2:2001 - Electromagnetic compatibility requirements and tests.
• EN1060-1: 1995 + Al: 2002 - Non-invasive sphygmomanometers- General
requirements .
• EN1060-3: 1997 + Al: 2005 Non-invasive sphygmomanometers supplementary
requirements for electromechanical blood pressure measuring systems.
• EN10604: 2004 - Non-invasive sphygmomanometers- Test procedures to
determine the overall system accuracy of automated non-invasive
sphygmomanometers.
Specifications
Method of measurement
Model number
Range of measurement
Accuracy
Inflation
Exhaust
Display
Sets of memory
Cuff size
Operating temperature
Storage temperature
Unit weight
Power supply
Battery life
Auto power off
Accessories
8
Oscillometric
BP4600
Pressure 0—300 mmHg
Pulse 40—199 beats/minute
Pressure +/- 3 mmHg
Pulse +/- 5% max.
Deluxe automatic
Automatic exhaust valve
Liquid crystal display — systolic, diastolic, pulse rate
32 sets
• Small/Medium cuff = 23—33 cm arm circumference
• Large/XLarge cuff = 33—43 cm arm circumference
+10 'C - +40 •C, less than R.H.
-20 -C - +60 •C, less than RH.
290g (without batteries)
Alkaline battery: 4 x AA (LR6)
300 times measurement (twice a day measurement)
Whenever not used for 1 minute
4 batteries, arm cuff with tube, instruction manual,
pouch, carton box
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC.
For detailed description Of EMC requirements please contact an authorized local Service
Centre (see package insert).
Portable and mobile RF communications equipment can affect medical electrical
equipment.
Please do not dispose of the product in the household waste at the end of its
useful life. Disposal can take place at your local retailer or at appropriate collection
points provided in your country.
Guarantee
See warranty statement inside the Kaz Customer Service Centre booklet.
UK only:
This guarantee in no way affects your rights under statutory law.
9