Microlife Tansiyon Aleti BP 3AS1-2 1. sayfa 1 Microlife Tansiyon Aleti BP 3AS1-2 2. sayfa 2 Microlife Tansiyon Aleti BP 3AS1-2 3. sayfa 3 Microlife Tansiyon Aleti BP 3AS1-2 4. sayfa 4 Microlife Tansiyon Aleti BP 3AS1-2 5. sayfa 5 Microlife Tansiyon Aleti BP 3AS1-2 6. sayfa 6 Microlife Tansiyon Aleti BP 3AS1-2 7. sayfa 7 Microlife Tansiyon Aleti BP 3AS1-2 8. sayfa 8 Microlife Tansiyon Aleti BP 3AS1-2 9. sayfa 9 Microlife Tansiyon Aleti BP 3AS1-2 10. sayfa 10 Microlife Tansiyon Aleti BP 3AS1-2 11. sayfa 11 Microlife Tansiyon Aleti BP 3AS1-2 12. sayfa 12 Microlife Tansiyon Aleti BP 3AS1-2 13. sayfa 13 Microlife Tansiyon Aleti BP 3AS1-2 14. sayfa 14 Microlife Tansiyon Aleti BP 3AS1-2 15. sayfa 15 Microlife Tansiyon Aleti BP 3AS1-2 16. sayfa 16


Microlife BP 3AS1-2 Tansiyon Aleti

M_3AS1_2_E&TR_rev2.qxd 09.04.2007 15:55 Page 10





   Other possible malfunctions and their elimination                                                       7. Care and maintenance, recalibration
   If problems occur when using the device, the following points should be checked and if necessary, the 
   corresponding measures are to be taken:                                                                 a) Do not expose the device to either extreme 
                                                                                                              temperatures, humidity, dust or direct sunlight.
   Malfunction                                      Remedy
   The display remains empty when the instrument is 1. Check whether the batteries are installed with
   switched on although the batteries are in place.   the correct polarity and correct if necessary.
                                                    2. If the display is unusual, remove the batteries and b) The cuff contains a sensitive air-tight bubble. Handle 
                                                      then exchange them for new.                             this carefully and avoid all types of straining through 
   The pressure does not rise although the          • Check the connection of the cuff tube and               twisting or buckling.
   pump is running.                                   connect properly if necessary. 
   The device frequently fails to measure the blood 1. Fit the cuff correctly on the arm.
   pressure values, or the values measured are too  2. Before starting the measurement make sure
   low (too high).                                    that the cuff is not fitted too tightly and that     c) Clean the device with a soft, dry cloth. Do not use 
                                                      there is no rolled-up sleeve exerting pressure on       petrol, thinners or similar solvent. Spots on the cuff can 
                                                      the arm above the measuring position.                   be removed carefully with a damp cloth and soapsuds. 
                                                      Take off articles of clothing if necessary.             The cuff must not be washed!
                                                    3. Measure the blood pressure again in complete 
                                                      peace and quiet.
   Every measurement results in different values,   • Please read the following information and
   although the device functions normally and normal  the points listed under «Common sources of           d) Handle the tube carefully. Do not pull on it. Likewise, 
   values are displayed.                              error». Repeat the measurement.                         this is not to be buckled or laid over sharp edges.
   The blood pressure values measured differ form   • Record the daily development of the values
   those measured by the doctor                       and consult your doctor about them.


                                                                                                           e) Do not drop the instrument or treat it roughly in any 
   t Further information                                                                                      way. Avoid strong vibrations.
   The level of blood-pressure is subject to fluctuations even with healthy people. Important thereby is, that 
   comparable measurements always require the same conditions (Quiet conditions)!
   If, in spite of observing all these factors, the fluctuations are larger than 15 mmHg, and/or you hear irregular 
   pulse tones on several occasions, please consult your doctor.                                           f) Never open the device! Otherwise the manufacturer 
                                                                                                              calibration becomes invalid!
   For licensing, the device has been subjected to strict clinical tests, by which the computer programme used to 
   measure the blood-pressure values was tested by experienced specialist doctors in Germany. The same 
   computer programme is used in every individual device, and has thus also been clinically tested.        Periodical recalibration
   The manufacture of the devices takes place according to the terms of the European standard for blood-   Sensitive measuring devices must from time to time be checked for accuracy. We therefore recommend a 
   pressure measuring devices (see technical data) under the supervision of the Technical Monitoring       periodical inspection of the static pressure display every 2 years.
   Association Essen (RWTüV-Essen).                                                                        Your specialist dealer would be pleased to provide more extensive information about this.

   You must consult your specialist dealer or chemist if there are technical problems with the blood-pressure 
   instrument. Never attempt to repair the instrument yourself!
   Any unauthorised opening of the instrument invalidates all guarantee claims!









   10                                                                                                                                                                                                  11