Microlife Tansiyon Aleti BP 3AS1-2 1. sayfa 1 Microlife Tansiyon Aleti BP 3AS1-2 2. sayfa 2 Microlife Tansiyon Aleti BP 3AS1-2 3. sayfa 3 Microlife Tansiyon Aleti BP 3AS1-2 4. sayfa 4 Microlife Tansiyon Aleti BP 3AS1-2 5. sayfa 5 Microlife Tansiyon Aleti BP 3AS1-2 6. sayfa 6 Microlife Tansiyon Aleti BP 3AS1-2 7. sayfa 7 Microlife Tansiyon Aleti BP 3AS1-2 8. sayfa 8 Microlife Tansiyon Aleti BP 3AS1-2 9. sayfa 9 Microlife Tansiyon Aleti BP 3AS1-2 10. sayfa 10 Microlife Tansiyon Aleti BP 3AS1-2 11. sayfa 11 Microlife Tansiyon Aleti BP 3AS1-2 12. sayfa 12 Microlife Tansiyon Aleti BP 3AS1-2 13. sayfa 13 Microlife Tansiyon Aleti BP 3AS1-2 14. sayfa 14 Microlife Tansiyon Aleti BP 3AS1-2 15. sayfa 15 Microlife Tansiyon Aleti BP 3AS1-2 16. sayfa 16


Microlife BP 3AS1-2 Tansiyon Aleti

M_3AS1_2_E&TR_rev2.qxd 09.04.2007 15:54 Page 8





     Inflate the cuff to a pressure at least 40 mmHg higher     5.6.Discontinuing a measurement
     than the expected systolic pressure. If you do not know    If it is necessary to interrupt a blood pressure measurement 
     the value of this pressure, inflate to a value of          for any reason (e.g. the patient feels
     approximately 160–180 mmHg. The cuff pressure is           unwell), press the quick release valve at the backside of the 
     shown continuously on the LCD.                             divece. The device then immediately lowers the cuff- 
                                                                pressure automatically.


   c) After the inflation process remain in a quiet and relaxed 5.7.Memory – displaying the last measurement 
     sitting position.                                          The blood-pressure monitor automatically stores the last 
     The measurement now proceeds on its own. If no             measurement value. By pressing the MEMORY button for 
     measurement takes place and all that appears in the        at least 3 seconds, the last measurement will be displayed.
     display is a flashing, upwards pointing arrow, see below 
     under «Inflate further». When the instrument detects a 
     pulse, a heart symbol in the display starts to flash and a 
     beep is heard for every heart beat.
                                                                t Further information
                                                                Measurements should not occur soon after each other, since otherwise, the results will be falsified. Wait 
   d) When the measurement has been concluded, a long           therefore for at least 1 minute in a relaxed position, sitting or lying, before you repeat a measurement.
     peep tone sounds. The measured systolic and diastolic 
     blood-pressure values as well as the pulse frequency 
     are now displayed.                                         6. Error messages/malfunctions
     The cuff pressure can now be quickly released by 
     pressing the release valve at the backside of the          If an error occurs during a measurement, the measurement 
     device.                                                    is discontinued and a corresponding error code is 
                                                                displayed. (example: error no. 1)

   5.5.Inflating further
   If the cuff is not inflated sufficiently, the measurement is 
   broken off after a few seconds and a flashing arrow pointing 
   upwards appears in the display.
   It is then necessary to inflate further to at least 20 mmHg  Error No. Possible cause(s) 
   higher than the previous value.                              ERR 1     The systolic pressure was determined but afterwards the cuff pressure fell below 20
   The instruction to inflate further can appear several times if         mmHg. This situation can occur, for example, if the tube becomes unplugged after the 
   the pressure is still not sufficient.                                  systolic blood pressure has been measured. Further possible cause: No pulse has been 
                                                                          detected 
   Attention:                                                   ERR 2     The pressure dropped below 20 mmHg without a measurement taking place. The 
   If the cuff is inflated too far (over 300 mmHg), «HI» appears          Release valve was pressed during the measurement. 
   in the display and a warning signal is sounded. The cuff     ERR 5     The measured readings indicated an unacceptable difference between systolic and 
   pressure must then be immediately released using the                   diastolic pressures. Take another reading following directions carefully. Contact your 
   quick-release valve) at the pump ball)!                                doctor if you continue to get unusual readings. 
   As soon as the cuff is inflated sufficiently and the 
   measurement is proceeding, continue as described under 
   points d–e.