Specifications
Method of measurement
Model number
Range of measurement
Accuracy
Inflation
Exhaust
Display
Sets of memory
Cuff size
Operating temperature
Storage temperature
Unit weight
Power supply
Battery life
Auto power off
Accessories
10
Oscillometric
BP4900, BP5900
Pressure 0—300 mmHg
Pulse 40—199 beats/minute
Pressure +/- 3 mmHg
Pulse 4/-5% Max.
Deluxe automatic
Automatic exhaust valve
Liquid crystal display — systolic, diastolic, pulse rate
BP4900: 48 sets
BP5900: 99 sets
Universal cuff 23—43 cm arm circumference
+10 -C +40 QC, less than RH
-20 oc- +60 -C, less than R.H.
Approximately 300g (without batteries)
Alkaline battery: 4 x AA (LR6)
Twice a day measurement:
BP4900: 300 times measurement
BP5900: 250 times measurement
Whenever not used for 1 minute
4 batteries, arm cuff with tube, instruction manual,
pouch, gift box
IMPORTANT
Read the operating instructions.
Classification:
• Internally powered equipment
• Type BF equipment
• Not suitable for use in the presence of flammable anesthetic mixture with air,
oxygen or nitrous oxide
• Continuous operation with short-time loading
Operating temperature
Storage temperature
This product conforms to the provisions of the EC directive 93/42/EEC (Medical
Device Directive). This device conforms to the following standards:
• EN 60601-1:1990 + Al: 1993 + A2: 1995 + Al 3: 1996 General requirementsforsafety.
• EN 60601-1-2:2001 - E ectromagnetic compatibility requirements and tests.
• EN1060-1: 1995 + Al: 2002 - Non-invasive sphygmomanometers- General
requirements .
• EN1060-3: 1997 + Al: 2005 Non-invasive sphygmomanometers supplementary
requirements for electromechanical blood pressure measuring systems.
• EN10604: 2004 - Non-invasive sphygmomanometers- Test procedures to
determine the overall system accuracy of automated non-invasive
sphygmomanometers.
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC.
For detailed description Of EMC requirements please contact an authorized local Service
Centre (see package insert).
Portable and mobile RF communications equipment can affect medical electrical
equipment.
Please do not dispose of the product in the household waste at the end of its
useful life. Disposal can take place at your local retailer or at appropriate collection
points provided in your country.
Guarantee
See warranty statement inside the Kaz Customer Service Centre booklet.
UK only:
This guarantee in no way affects your rights under statutory law.
11
İlginizi çekebilecek Braun marka Tansiyon Aleti ürünleri