NIBP Monitoring ■ NIBP alarm setups ALM: pick "ON", the system will give alarm prompt and store alarm information once NIBP alarm occurs; pick "OFF", the system will not give any alarm information and instead display a & beside "NIBP". ALM LEV: selectable from HIGH, MED to LOW. HIGH represents the most serious case. ALM REC: pick "ON", the system will command the recorder to output the alarm information once NIBP alarm occurs. SYS ALM HI, SYS ALM LO, MEAN ALM HI, MEAN ALM LO, DIA ALM HI, DIA ALM LO are for you to set up the alarm limit for each type of pressure. NIBP alarm is activated when the pressure exceeds the upper alarm limit or falls below lower alarm limit. NIBP alarm limits: Adult Mode SYS 40-270 mmHg DIA 10-215 mmHg Mean 20-235 mmHg Pediatric Mode SYS 40-200 mmHg DIA 10-150 mmHg Mean 20-165 mmHg Neonatal Mode SYS 40-135 mmHg DIA 10-100 mmHg Mean 20-110 mmHg ■ UNIT Options are mmHg and kPa. ■ INTERVAL Time interval for AUTO measurement. Available selections are: 1/2/3/4/5/10/15/30/60/90/120/180/240/480 minutes. After selecting the INTERVAL, the system will display the prompt "Please press START button" in the NIBP prompt area. At this time, you can press the NIBP button on the front panel to start the first AUTO measurement. Pick the [MANUAL] item in the list of INTERVAL and the system will end AUTO measurement and restore MANUAL measurement mode. ■ INFLATION Pick this item to choose the initial pressure when inflate the cuff next time. There are different selections of the initial pressure. Default Default initial Options for Initial value in NIBP MANUAL value menu (mmHg/kPa) (mmHg/kPa) FACTORY DEFAULT 150 80/100/120/140/150/160/180/200/220/240 ADU CONFIG FACTORY DEFAULT 100 80/100/120/140/150/160/180/200 PED CONFIG 13-127