NIBP Monitoring
If the patient is in severe shock or hypothermia, measurements will be unreliable since reduced
blood flow to the peripheries will cause reduced pulsation of the arteries.
• Heart Rate Extremes
Measurements can not be made at a heart rate of less than 40 bpm and greater than 240 bpm.
• Fat Patient
The thick fat under extremity may reduce the measurement accuracy because it may prevent
the oscillation generated in the artery from reaching the cuff.
13.2.2 NIBP monitoring screen
NIBP measurement result and corresponding message are displayed as follows:
~^ Measurement
Measurement value time - BP unit mmHg
Measurement mode— NS Alarm limit 70 160 90
/— Current cuff
pressure CUFF:100