NIBP Monitoring
FACTORYDEFAULTNEO 70 60/70/80/100/120
CONFIG
USER DEFAULT ADU 150 80/100/120/140/150/160/180/200/220/240
CONFIG
USER DEFAULT PED 100 80/100/120/140/150/160/180/200
CONFIG
USER DEFAULT NEO 70 60/70/80/100/120
CONFIG
Press the MENU hot key, the system enters the DEAULT menu of SYSTEM MENU, in
which you can select one of the six available configurations. After confirming your
selection, return to the main screen and pick the NIBP hot key in the NIBP area on the
main screen, the system will access the NIBP SETUP menu, in which the value shown in
the [INFLATION] item is the initial inflating pressure value corresponding to the selected
default configuration. Highlight the [INFALTION] item and pick it, you can see the initial
inflation range as shown above.
A A
Warning
The [INFLATION] item is to help you to select the initial inflation pressure of the cuff for
the next time. However, the afterward initial pressure during measurement is based on
the previous measured SYS value of the same patient. The system will remember this
value so that the time required to measure the same patient can be reduced and also
the accuracy increased.
A Warning A
If you only make setup in the [PATIENT TYPE] item of the PATIENT SETUP menu but
not choose any option in DEFAULT, the system will automatically make initial setups
for the relevant modules as per PTIENT TYPE. Also, any change in the DEFAULT may
cause the PATIENT TYPE in PATIENT SETUP to change accordingly.
■ RESET
Restore measurement status of NIBP pump.
Pick this item to restore initial settings of the pressure pump.
When the pressure does not work properly and the system fails to give message for the
problem, pick this item to activate self-test procedure, thus restore the system from
abnormal performance.
■ CONTINUAL
Start CONTINUOUS measurement.
When this item is picked, the menu will disappear automatically and the CONTINUOUS
measurement start immediately.
■ CALIBRATE
It is recommended to use a calibrated pressure meter with the precision higher than
1mmHg to calibrate the device. Pick the [CALIBRATE] item to start the calibration
process and this item will display "STOP CAL". If pick this item at this time, the system
will stop the calibration.
1^