NIBP Monitoring FACTORYDEFAULTNEO 70 60/70/80/100/120 CONFIG USER DEFAULT ADU 150 80/100/120/140/150/160/180/200/220/240 CONFIG USER DEFAULT PED 100 80/100/120/140/150/160/180/200 CONFIG USER DEFAULT NEO 70 60/70/80/100/120 CONFIG Press the MENU hot key, the system enters the DEAULT menu of SYSTEM MENU, in which you can select one of the six available configurations. After confirming your selection, return to the main screen and pick the NIBP hot key in the NIBP area on the main screen, the system will access the NIBP SETUP menu, in which the value shown in the [INFLATION] item is the initial inflating pressure value corresponding to the selected default configuration. Highlight the [INFALTION] item and pick it, you can see the initial inflation range as shown above. A A Warning The [INFLATION] item is to help you to select the initial inflation pressure of the cuff for the next time. However, the afterward initial pressure during measurement is based on the previous measured SYS value of the same patient. The system will remember this value so that the time required to measure the same patient can be reduced and also the accuracy increased. A Warning A If you only make setup in the [PATIENT TYPE] item of the PATIENT SETUP menu but not choose any option in DEFAULT, the system will automatically make initial setups for the relevant modules as per PTIENT TYPE. Also, any change in the DEFAULT may cause the PATIENT TYPE in PATIENT SETUP to change accordingly. ■ RESET Restore measurement status of NIBP pump. Pick this item to restore initial settings of the pressure pump. When the pressure does not work properly and the system fails to give message for the problem, pick this item to activate self-test procedure, thus restore the system from abnormal performance. ■ CONTINUAL Start CONTINUOUS measurement. When this item is picked, the menu will disappear automatically and the CONTINUOUS measurement start immediately. ■ CALIBRATE It is recommended to use a calibrated pressure meter with the precision higher than 1mmHg to calibrate the device. Pick the [CALIBRATE] item to start the calibration process and this item will display "STOP CAL". If pick this item at this time, the system will stop the calibration. 1^