Mindray MEC-1200 Hastabaşı Monitörü
ECG/RESP Monitoring • DEFAULT: pick this item to access the RESP DEFAULT CONFIG dialog box, in which you can select FACTORY DEFAULT CONFIG or USER DEFAULT CONFIG. After selecting an item and exiting the dialog box, the system will pop up the dialog box asking for the user's confirmation 11.9.5 RESP Alarm Information and Prompt Information If the alarm recording switches in relevant menus are set to ON, the physiological alarms caused by parameter exceeding alarm limit can trigger the recorder to automatically output alarming parameter value and corresponding measuring waveform. Tables below list out the possible physiological alarms, technical alarms and prompt messages during RESP measurement. Physiological alarms: Message Cause Alarm Level RR TOO HIGH RESP measuring value is above User-selectable upper alarm limit. RR TOO LOW RESP measuring value is below User-selectable lower alarm limit. RESP APNEA RESP can not be measured within HIGH specific time interval. Technical alarms: Message Cause Alarm Level Remedy RESP ALM LMT ERR Functional safetyHIGH Stop using RESP failure alarming function, notify biomedical engineer or Mindray service staff. Prompt message (general alarms): Message Cause Alarm Level RR EXCEED RR measuring value exceeds the HIGH measure range. 11.10 Maintenance and Cleaning Care and Cleaning A Warning A Turn off the power and disconnect the line power before cleaning the monitor or the 11-108