Microlife Tansiyon Aleti BP 3BU1-3 1. sayfa 1 Microlife Tansiyon Aleti BP 3BU1-3 2. sayfa 2 Microlife Tansiyon Aleti BP 3BU1-3 3. sayfa 3 Microlife Tansiyon Aleti BP 3BU1-3 4. sayfa 4 Microlife Tansiyon Aleti BP 3BU1-3 5. sayfa 5 Microlife Tansiyon Aleti BP 3BU1-3 6. sayfa 6 Microlife Tansiyon Aleti BP 3BU1-3 7. sayfa 7 Microlife Tansiyon Aleti BP 3BU1-3 8. sayfa 8 Microlife Tansiyon Aleti BP 3BU1-3 9. sayfa 9 Microlife Tansiyon Aleti BP 3BU1-3 10. sayfa 10 Microlife Tansiyon Aleti BP 3BU1-3 11. sayfa 11 Microlife Tansiyon Aleti BP 3BU1-3 12. sayfa 12 Microlife Tansiyon Aleti BP 3BU1-3 13. sayfa 13 Microlife Tansiyon Aleti BP 3BU1-3 14. sayfa 14 Microlife Tansiyon Aleti BP 3BU1-3 15. sayfa 15 Microlife Tansiyon Aleti BP 3BU1-3 16. sayfa 16 Microlife Tansiyon Aleti BP 3BU1-3 17. sayfa 17 Microlife Tansiyon Aleti BP 3BU1-3 18. sayfa 18 Microlife Tansiyon Aleti BP 3BU1-3 19. sayfa 19 Microlife Tansiyon Aleti BP 3BU1-3 20. sayfa 20 Microlife Tansiyon Aleti BP 3BU1-3 21. sayfa 21 Microlife Tansiyon Aleti BP 3BU1-3 22. sayfa 22


Microlife BP 3BU1-3 Tansiyon Aleti

M-BP 3BU1-3_E&TR.qxd 09.11.2004 10:47 Page 14




    5.6. Memory – storage and recall of the measurements                                                 Other possible malfunctions and their elimination
                                                                                                         If problems occur when using the device, the following points should be checked and if necessary, 
    The blood-pressure monitor automatically stores each of                                              the corresponding measures are to be taken:
    the last 30 measurement values. By pressing the 
    MEMORY button, the last measurement (MR1) as well as                                                 Malfunction                                    Remedy
    the further last 29 measurements (MR2, MR3, ..., MR30)                                               The display remains empty when the             1. Check batteries for correct polarity and if
    can be displayed one after the other.                                                                instrument is switched on although the           necessary insert correctly
                                                                                                         batteries are in place.                        2. If the display is unusual, re-insert batteries or
                                                                                                                                                          exchange them. 
                                                                                                         The device frequently fails to measure the     1. Check the positioning of the cuff.
                                                                                                         blood pressure values, or the values measured 2. Measure the blood-pressure again in peace
                                                                                                         are too low (too high).                          and quiet under observance of the details
                                                                                                                                                          made under point 5. 
    (MR1: Values of the last measurement)   (MR2-MR30: Values of the measurement before MR1)             Every measurement results in different values, 1. Please pay attention to the following
                                                                                                         although the device functions normally and       information as well as the points listed
    ~ Further information                                                                                normal values are displayed.                     under «Common sources of error» and repeat
    Measurements should not occur soon after each other, since otherwise, the results will be falsified.                                                  the measurement.
    Wait therefore for several minutes in a relaxed position, sitting or lying, before you repeat a      The blood pressure values measured differ      1. Record the daily development of the values
    measurement.                                                                                         from those measured by the doctor                and consult your doctor about them 

    6. Error messages / malfunctions                                                                     ~ Further Information
                                                                                                         The level of blood-pressure is subject to fluctuations even with healthy people. Important 
    If an error occurs during a measurement, the                                                         thereby is, that comparable measurements always require the same conditions (Quiet 
    measurement is discontinued and a corresponding error                                                conditions)!
    code is displayed (Example: Error No. 2).                                                            If, in spite of observing all these factors, the fluctuations are larger than 15mmHg, and/or you hear 
                                                                                                         irregular pulse tones on several occasions, please consult your doctor.
    Error No.    Possible cause(s)
    Err 1        No pulse has been detected.
    Err 2        Unnatural pressure impulses influence the measurement result. Reason: The arm 
                 was moved during the Measurement (Artefact).
    Err 3        The inflation of the cuff takes too long. The cuff is not correctly seated.
    14                                                                                                                                                                                             15