ECG/RESP Monitoring generate alarming information. Arrhythmia algorithm can monitor paced and non-paced patients. Qualified personnel can use arrhythmia analysis to evaluate patient's condition (such as heart rate, PVCs frequency, rhythm and ectopic beat) and decide the treatment. Besides detecting changing of ECG, arrhythmia analysis can also monitor patients and give proper alarm for arrhythmia. ■ The arrhythmia monitoring is shutoff by default. You can enable it when necessary. This function can raise the doctor's attention to the patient's heart rate by measuring and classifying the arrhythmia and abnormal heart beat and triggering the alarm. ■ The monitor can conduct up to 13 different arrhythmia analyses. The monitor can store the latest 60 alarm events (4 seconds prior to and after alarm). You can edit these arrhythmia events through using the menu below. Pick the [ARR ANALYSIS] item in the ECG SETUP menu to access the ARR ANALYSIS menu. ARR ANALYSIS Menu Figure 11-11 ARR ANALYSIS Menu ARR ANAL: Pick "ON" during monitoring. Default set is "OFF". PVCs ALM: Pick "ON" and the system will give and store alarm information when alarm is occurring. Pick "OFF" and the system will not give PVCs alarm, instead, it will display aife beside "PVCs". ALM LEV: there are three options: HIGH, MED, LOW. Level HIGH represents the most serious PVCs alarm. ALM REC: pick "ON" and the system will command the recorder to output information when PVCs alarm is occurring. ALM HI: the system will give PVCs alarm once PVCs exceed the set PVCs ALM HI value. 11-99