ECG/RESP Monitoring
generate alarming information. Arrhythmia algorithm can monitor paced and non-paced
patients. Qualified personnel can use arrhythmia analysis to evaluate patient's condition (such
as heart rate, PVCs frequency, rhythm and ectopic beat) and decide the treatment. Besides
detecting changing of ECG, arrhythmia analysis can also monitor patients and give proper
alarm for arrhythmia.
■ The arrhythmia monitoring is shutoff by default. You can enable it when necessary.
This function can raise the doctor's attention to the patient's heart rate by measuring and
classifying the arrhythmia and abnormal heart beat and triggering the alarm.
■ The monitor can conduct up to 13 different arrhythmia analyses.
The monitor can store the latest 60 alarm events (4 seconds prior to and after alarm).
You can edit these arrhythmia events through using the menu below.
Pick the [ARR ANALYSIS] item in the ECG SETUP menu to access the ARR ANALYSIS
menu.
ARR ANALYSIS Menu
Figure 11-11 ARR ANALYSIS Menu
ARR ANAL: Pick "ON" during monitoring. Default set is "OFF".
PVCs ALM: Pick "ON" and the system will give and store alarm information when alarm
is occurring. Pick "OFF" and the system will not give PVCs alarm, instead, it will display
aife beside "PVCs".
ALM LEV: there are three options: HIGH, MED, LOW. Level HIGH represents the most
serious PVCs alarm.
ALM REC: pick "ON" and the system will command the recorder to output information
when PVCs alarm is occurring.
ALM HI: the system will give PVCs alarm once PVCs exceed the set PVCs ALM HI
value.
11-99