Kullanım Kılavuzları
ECG/RESP Monitoring established, the system will display message "ST analysis key off!"). You can adjust the position of the highlighted line in the window. The method is first select ISO or ST, then turn the knob clockwise or counterclockwise to move the highlighted line horizontally so as to define the reference point or the measurement point. ZlA Note ^-lA Abnormal QRS complex is not considered in ST segment analysis. ST Alarm Message If the alarm recording switches in relevant menus are set to ON, the physiological alarms caused by parameter exceeding alarm limit can trigger the recorder to automatically output alarming parameter value and corresponding measuring waveform. Tables below list out possible physiological alarms, technical alarms and prompt messages that may occur during ST measurement. Physiological alarms: Message Cause Alarm Level ST TOO HIGH ST measuring value is above the upper User-selectable alarm limit. ST TOO LOW ST measuring value is below the lower User-selectable alarm limit. Technical alarms: Message Cause Alarm Level Remedy ST ALM LMT ERR Functional safety failure HIGH Stop using ST alarming function, notify biomedical engineer or Mindray service staff. Prompt messages (include general alerts): Message Cause Alarm Level ST EXCEED ST measuring value exceeds the measurement HIGH range. 11.8 Arr. Monitoring (optional) Arrhythmia Analysis The arrhythmia algorithm is used to monitor ECG of neonate and adult patient in clinical, detect the changing of heart rate and ventricular rhythm, and also save arrhythmia events and 11-98