NIBP Monitoring
Warning A
You shall calibrate NIBP measurement once every two years (or as required in your
hospital's maintenance regulation). You shall check the performance according to
following information.
Procedures to calibrate Pressure Transducer:
Replace the cuff of the device with a rigid metal vessel with a capacity of 500 ml t 5%.
Connect a calibrated reference manometer with an error less than 0.8 mmHg and a ball pump
by means of a T-piece connector and hoses to the pneumatic system. Set the monitor in
CALIBRATE mode. Inflate the pneumatic system to 0, 50 and 200 mmHg by ball pump
separately. The difference between the indicated pressure of the reference manometer and
the indicated pressure of the monitor will not exceed 3 mmHg. Otherwise, please contact our
customer service.
Monitor
NIBP cuff
connector
Ball pump
Figure 13-3 Diagram of NIBP calibration
■ PNEUMATIC
This item is used for air leakage test. After connecting NIBP cuff, you can pick this item to
start NIBP inflating process so as to find if the NIBP airway is closed and in good
condition. If the test is successful, the system will not give any prompt. If the test is failed,
the system will display corresponding error message in the NIBP information area.
■ DEFAULT
Pick this item to enter the NIBP DEFAULT dialog box, in which you can choose factory
configuration or user configuration. After choosing one configuration, the system will
display a confirmation box for you to confirm your selection.
â\ Warning A
This pneumatic test other than being specified in the EN 1060-1 standard is to be used
by you to simply determine whether there are air leaks in the NIBP airway. If at the end
of the test the system gives the prompt that the NIBP airway has air leaks, please
contact the manufacturer for repair.