NIBP Monitoring Warning A You shall calibrate NIBP measurement once every two years (or as required in your hospital's maintenance regulation). You shall check the performance according to following information. Procedures to calibrate Pressure Transducer: Replace the cuff of the device with a rigid metal vessel with a capacity of 500 ml t 5%. Connect a calibrated reference manometer with an error less than 0.8 mmHg and a ball pump by means of a T-piece connector and hoses to the pneumatic system. Set the monitor in CALIBRATE mode. Inflate the pneumatic system to 0, 50 and 200 mmHg by ball pump separately. The difference between the indicated pressure of the reference manometer and the indicated pressure of the monitor will not exceed 3 mmHg. Otherwise, please contact our customer service. Monitor NIBP cuff connector Ball pump Figure 13-3 Diagram of NIBP calibration ■ PNEUMATIC This item is used for air leakage test. After connecting NIBP cuff, you can pick this item to start NIBP inflating process so as to find if the NIBP airway is closed and in good condition. If the test is successful, the system will not give any prompt. If the test is failed, the system will display corresponding error message in the NIBP information area. ■ DEFAULT Pick this item to enter the NIBP DEFAULT dialog box, in which you can choose factory configuration or user configuration. After choosing one configuration, the system will display a confirmation box for you to confirm your selection. â\ Warning A This pneumatic test other than being specified in the EN 1060-1 standard is to be used by you to simply determine whether there are air leaks in the NIBP airway. If at the end of the test the system gives the prompt that the NIBP airway has air leaks, please contact the manufacturer for repair. 13-129