SpO2 Monitoring
Physiological alarm:
Message Cause Alarm Level
SPO2 TOO HIGH SpO2 measuring value is above upper alarm limit. User-selectable
SpO2 TOO LOW SpO2 measuring value is below lower alarm limit. User-selectable
PR TOO HIGH PR measuring value is above upper alarm limit. User-selectable
PR TOO LOW PR measuring value is below lower alarm limit. User-selectable
Message Cause Alarm Remedy
Level
SPO2 SENSOR OFF SpO2 sensor may LOW Make sure that the
be disconnected monitor and the patient
from the patient or are in correct connection
the monitor. with the cables.
SpO2 module HIGH Stop using the measuring
SPO2 INIT ERR
failure function of SpO2 module,
SPO2 INIT ERR 1 notify biomedical
engineer or Mindray
SPO2 INIT ERR 2 service staff.
SPO2 INIT ERR 3
SPO2 INIT ERR 4
SPO2 INIT ERR 5
SPO2 INIT ERR 6
SPO2 INIT ERR 7
SPO2 INIT ERR 8
SPO2 COMM STOP SpO2 module HIGH Stop using the measuring
failure or function of SpO2 module,
communication error notify biomedical
engineer or Mindray
service staff.
SPO2 COMM ERR SpO2 module HIGH Stop using the measuring
failure or function of SpO2 module,
communication error notify biomedical
engineer or Mindray
service staff.
SPO2 ALM LMT ERR Functional safety HIGH Stop using the measuring
failure function of SpO2 module,
notify biomedical
engineer or Mindray
service staff.
PR ALM LMT ERR Functional safety HIGH Stop using the measuring
failure function of SpO2 module,
notify biomedical
engineer or Mindray
service staff.
12-118