SpO2 Monitoring Physiological alarm: Message Cause Alarm Level SPO2 TOO HIGH SpO2 measuring value is above upper alarm limit. User-selectable SpO2 TOO LOW SpO2 measuring value is below lower alarm limit. User-selectable PR TOO HIGH PR measuring value is above upper alarm limit. User-selectable PR TOO LOW PR measuring value is below lower alarm limit. User-selectable Message Cause Alarm Remedy Level SPO2 SENSOR OFF SpO2 sensor may LOW Make sure that the be disconnected monitor and the patient from the patient or are in correct connection the monitor. with the cables. SpO2 module HIGH Stop using the measuring SPO2 INIT ERR failure function of SpO2 module, SPO2 INIT ERR 1 notify biomedical engineer or Mindray SPO2 INIT ERR 2 service staff. SPO2 INIT ERR 3 SPO2 INIT ERR 4 SPO2 INIT ERR 5 SPO2 INIT ERR 6 SPO2 INIT ERR 7 SPO2 INIT ERR 8 SPO2 COMM STOP SpO2 module HIGH Stop using the measuring failure or function of SpO2 module, communication error notify biomedical engineer or Mindray service staff. SPO2 COMM ERR SpO2 module HIGH Stop using the measuring failure or function of SpO2 module, communication error notify biomedical engineer or Mindray service staff. SPO2 ALM LMT ERR Functional safety HIGH Stop using the measuring failure function of SpO2 module, notify biomedical engineer or Mindray service staff. PR ALM LMT ERR Functional safety HIGH Stop using the measuring failure function of SpO2 module, notify biomedical engineer or Mindray service staff. 12-118