SpO2 Monitoring
Prompt message (include general alerts):
Message Cause Alarm Level
SPO2 EXCEED SpO2 measuring value exceeds the range. HIGH
PR EXCEED PR measuring value exceeds the range. HIGH
SEARCH PULSE SpO2 module is searching for pulse. No alarm
NO PULSE SpO2 module cannot detect SpO2 signal for a long HIGH
time.
12.7 Maintenance and Cleaning
Care and Cleaning
Warning A
Turn of the monitor and disconnect the line power before cleaning the monitor or the
sensor
Warning A
Do not subject the sensor to autoclaving.
Do not immerse the sensor into any liquid.
Do not use any sensor or cable that may be damaged or deteriorated.
Cleaning:
Use a cotton ball or a soft mull moistened with hospital-grade ethanol to wipe the surface
of the sensor, and then dry it with a cloth. This cleaning method can also be applied to
the luminotron and receiving unit.
The cable can be cleaned with 3% hydrogen dioxide, 7% isopropanol, or other active
reagent. However, connector of the sensor shall not be subjected to such solution.