SpO2 Monitoring SPO2 ALM HI value or falls below SPO2 ALM LO value. ■ PR ALM HI and PR ALM LO: PR alarm is activated when the pulse rate exceeds set PR ALM HI value or falls below PR ALM LO value. SpO2 and PR alarm limits: Max. Upper Limit Min. Lower Limit Step SpO2 100 0 1 PR 254 0 1 The default SpO2 and PR alarm limits: Parameters Max. Upper Limit Min. Lower Limit SpO2 Adult 100 90 Pediatric 100 90 Neonatal 95 80 PR Adult 120 50 Pediatric 160 75 Neonatal 200 100 ■ SWEEP Available options are 12.5mm/s, 25.0 mm/s. ■ PR SOUND Pulse beep volume. Options are OFF, HIGH, MED, LO W. ■ AVG TIM E 4s, 8s, 16s represent times that SpO2 average value is counted. ■ DEFAULT Pick this item to access the SPO2 DEFAULT CONFIG dialog box, in which you can select FACTORY DEFAULT CONFIG or USER DEFAULT CONFIG. After selecting one item and exiting the dialog box, the system will pop up the dialog box asking for the your confirmation. 12.6 Alarm Description and Prompt SpO2 Alarm Message When the alarm switches are set to ON in relevant menus, the physiological alarms caused by the parameter exceeding the alarm limit may possibly trigger the recorder to automatically output alarming parameter value and corresponding waveforms. Tables below describe the possible physiological alarms, technical alarms and prompt messages occurring during SpO2 measurement. 12-117