SpO2 Monitoring (carbon hemoglobin, methemoglobin, methylene blue and indigo carmine) absorbing the light of the same wavelength, they may result in false or low SpO2 value. • It is recommended to use SpO2 sensors described in chapter Accessories and Ordering Information. 12.5 SpO2 Menu SPO2 SETUP Menu Turn the knob to move the cursor onto the SPO2 hot key in the Parameter area, push the knob to access the SPO2 SETUP menu. srui: SETUF ALM ] PH ALU LU OX ft[.H LEU 58 HEB 5UEEP Z5HQ flLfl TlEC OFF M SOUND nED SP02 ALU HI 100 AUG riflE SFOZ ALM LD 30 DEFAULT » FR ALU HI 120 Dpen or close the SpDZ iilarn. EXIT Figure 12-5 SPO2 SETUP menu A Warning A Setting the SpO2 upper alarm limit to 100% is equivalent to switching off the alarm on upper limit. High oxygen levels may predispose a premature infant to retrolental fibroplasia. Therefore, the upper alarm limit for oxygen saturation must be carefully selected in accordance with commonly accepted clinical practices. SpO2 alarm setting ALM: pick "ON", the system will give alarm prompt and store alarm information when SpO2 alarm occurs; pick "OFF", the system will not give alarm and instead display a ife beside "SpO2". ALM REC: pick "ON", the system will command the recorder to output alarm information when SpO2 alarm occurs. ALM LEV: used to set up alarm level, selectable from HIGH, MED and LOW. HIGH represents the most serious case. SPO2 ALM HI and SPO2 ALM LO: SpO2 alarm is activated when the result exceeds set 12-116