Kullanım Kılavuzları
Tips for Measuring at Different Sites Under-the-arm (axillary) use This method is used for babies and young children. I. Wipe the underarm witha dry towel. 2. Place the probe tip under the arm so the tip is touching the skin and position the patients arm next to the patient% body (Fig With a young child, it is sometimes helpful to hug the child to keep their arm next to their body. This ensures that the room air does not affect the reading, When the peak temperature is reached, beep Signals Will is confirmed. When thermometer is removed from the measuring site, the associated colour will display. 4 Turn offthe thermometer and clean as recommended below. Fig. 1 Oral use Fig. 2 3. Rotate battery door to LOCKED position. Errors and troubleshooting When measured temperature is hiaher than 42_9 LCD Will display "Err •C" and relevant age icon. Red backlight stays On for 5 seconds. ambient temperature is out Of range product Specifications below) or other system errors occur such as short or bmken circuit, LCD will display•Err" to indicate system error. Red backlight on for S seconds. Calibration This device Was designed and manufactured for a long service life, it is recommended to have the device inspected once a year by the Authorized Servite Center located in your to ensure Correct function and Checking is not a free service and therefore We recommend that you Contact the Authorized Service Centerto get a quotation before you send out the product. This method is suitable for children who are old enough to safety keep the thermometer inside the mouth. I. It's important to place the tip ofthethermometer well under the tongue on either side to cet a good, accurate reading. Place sensor tip in areas marked with the tongue (Fig. 2)_ keep patient's mouth closed and make him/her sit still to help ensure an accurate measurement. 2. When the peak temperature is reached, beep Signals v.'"l sound. The temperature is now confirmed. When thermometer is removed from the measuring site, the associated Will display. 3, Turn Offthe thermometer and Clean as recommended below, Rectal use Commonly used for babies, young children, or when it is difficult to take an oral or underarm temperature. Lubricate the thermometer'S tip With a for easier insertion, Do not use a petroleum jelly. Read the label of the lubricant to be sure you are using the right kind. 2, Lay the patient on his/her Side. "the patient is an infant, the proper position fora baby is to lay on its stomach with legs hanginQ down, either across your knees or at the edge of a bed or Changing table. This positions the rectum for safe and easy insertion Of the thermometer. 3. Gently insert the tip ofthc thermometer NO MORE THAN cm into the rectum. Ifyou detect resistance, Stop, Hold the thermometer in place during measurement, 4. When the peak temperature is reached, beep Signals Will sound, The temperature is now confirmed. When thermometer is removed from the measuring site, the associated colour will display. 5. Turn offthe thermometer and dean as recommended below. Memory When the thermometer is turned on, the display will show a test, after which the last measured temperature will be shown for approximately second Letter •M"will be displayed next to the temperature reading, Changing the Temperature Scale With power off, press and hold the "On,'Off' button for about 5 seconds until"- •v and a flashing "t • Or appear on screen. Release and press the •On/Off" button again to Switch from the current scale to the alternate sca'e and release. After about 3 seconds the unit is ready for taking temperature in the Chosen scale and the age range Start to scroll. Care and Cleaning Clean the probe tip by wash with soap and warm water or by disinfecting with rubbing alcohol (70% isopropyl a Wipe dry with soft cloth. DO NOT BOIL OR CLEAN IN DISHWASHER. This will cause the thermometer to no longer function and will void the DO not wash the unit With any thinner or Chemical Store Vour thermometer in a dry location, free from dust and contaminants and away from direct sunlight, Replacing the Batteries When the battery symbol on LCD display twinkles continually @.please replace witha new batte I. Insert coin into slot. Rotate door until slot is lined with UNLOCK position. 2. Replace battery. Product specifications Displayed temperature range: Operating ambient temperature range: Display resolution: P«uracy for displayed temperature range 35.5 oc-42 outside this range: Battery type: Battery life: Service life Storage/Transport temperature and relative humidity 32.0 'F - 109.2 'F) 10-40 oc 104 •F) and relative humidity range of 0.1 acorns Maximum Laboratory Error 1±0.2 ambientof23±5 'C ±0.2 •C,'0.4 (within 32.0 •C-42.9 oc -1092 OF)) W DC, IXCR1637 51K' times measurements within 2 years 2000 measurements -25-55 'C ± 2 -Cand 15-95 If the device is not used within specified temperature and humidity ranges the technical accuracy of the measurement Cannot be guaranteed, Explanation Of symbols Type BF applied parts Operating temperature Keep dry Internally powered equipment. Continuous operation. Not intended to be sterilised. Not for use in an rich environment. See for use Storage tem re IP22: Protected against solid fore.gn Objects Of 12 S mm diameter and greater, protected against vertically falling water drops when the device is tilted up to 1 5 degrees. This thermometer is specified to operate at 1 atmospheric pressure or at altitudes With an atmospheric pressure up to atmospheric pressure (160-1060 hPa). Equipment with type BF applied parts. Subject to change without notice. This appliance conforms to the following Standards: Standard Reference Edition EN 1 Clinical thermometers —part performance Of compact electrical thermometers (non-predictive and predictive with maximum device EN 60601- 2006 Medical electrical equipment— Part I: General requirements for basic safety and essential performance. EN ISO 14971: 2012 Medical devices - Application of risk management to medical devices. EN ISO 10993-1 2009 Biological evaluation of medical devices — Part Evaluation and Testing. EN 60601-1-2:2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance — Collateral standard: electromagnetic compatibility - Requirements and tests, EN Symbols for use in labeling Of medical devices, EN 1041 2008 Information supplied by the manufacturer of medical devices. EN 60601- I-I I: 201 C Medical electrical equipment - Part I-Il: General requirements for basic safety and essential performance — Collateral Standad Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. This product conforms to the provisions of the FC directive 93/42/FFC. MEDICAL ELECTRICAL EQUIPMENT needs special Eye-cautions regarding EMC, detailed description of EMC requirements please contact •,our authorized local Service Centre (See guarantee card]. Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.