Kullanım Kılavuzları
ECG/RESP Monitoring actual waveform so as to enable the system to display complete waveform. @ Filter method: used to obtain clearer and more accurate waveform There are three selectable filter modes. In DIAGNOSTIC mode, the system displays the ECG waveform that has not been filtered. In MONITOR mode, the artifacts that may lead to false alarm can be filtered. In SURGERY mode, the artifacts and interference from the ES equipment can be reduced. The system displays the filter mode above the ECG waveform. A Warning The system displays non-processed real signal only in Diagnostic mode. In Monitor or Surgery mode, ECG waveforms may have distortion of different extent. In either of the latter two modes, the system can only show the basic ECG and the results of ST analysis may also be greatly affected. In Surgery mode, ARR analysis result may be somewhat affected. Therefore, it is suggested that in the environment having relative small interference, you'd better monitor a patient in Diagnostic mode. Z!a Note ^-lA The system will indicate the detected pacemaking signal by displaying above the ECG waveform. 11.5 ECG Menu ECG SETUP Menu Turn the knob to highlight the ECG hot key in the Parameter Area on the main screen and then push the knob to call up the ECG SETUP menu. Figure 11-4 ECG SETUP menu 11-90