ECG/RESP Monitoring
actual waveform so as to enable the system to display complete waveform.
@ Filter method: used to obtain clearer and more accurate waveform There are three
selectable filter modes. In DIAGNOSTIC mode, the system displays the ECG waveform that
has not been filtered. In MONITOR mode, the artifacts that may lead to false alarm can be
filtered. In SURGERY mode, the artifacts and interference from the ES equipment can be
reduced. The system displays the filter mode above the ECG waveform.
A
Warning
The system displays non-processed real signal only in Diagnostic mode. In Monitor or
Surgery mode, ECG waveforms may have distortion of different extent. In either of the
latter two modes, the system can only show the basic ECG and the results of ST
analysis may also be greatly affected. In Surgery mode, ARR analysis result may be
somewhat affected. Therefore, it is suggested that in the environment having relative
small interference, you'd better monitor a patient in Diagnostic mode.
Z!a Note ^-lA
The system will indicate the detected pacemaking signal by displaying above the
ECG waveform.
11.5 ECG Menu
ECG SETUP Menu
Turn the knob to highlight the ECG hot key in the Parameter Area on the main screen and
then push the knob to call up the ECG SETUP menu.
Figure 11-4 ECG SETUP menu
11-90