Kullanım Kılavuzları
NIBP Monitoring NM TOO LOW NIBP MAP measuring value is User-selectable below lower alarm limit. _______ Technical alarms 1: (display in Information area) Message Cause Alarm Remedy Level NS ALM LMT Functional safety failure HIGH Stop using NIBP alarm function and ERR notify biomedical engineer or Mindray service staff. NM ALM LMT Functional safety failure HIGH Stop using NIBP alarm function and ERR notify biomedical engineer or Mindray service staff. ND ALM LMT Functional safety failure HIGH Stop using NIBP alarm function and ERR notify biomedical engineer or Mindray service staff. Technical alarms 2: (display in the area below the NIBP value) Message Cause Alarm Remedy Level NIBP SELFTEST Sensor or other HIGH Stop using NIBP measurement ERR hardware of NIBP function, notify biomedical engineer or module is incorrect. Mindray service staff. NIBP COMM ERR Communication with HIGH If failure persists, NIBP measurement, NIBP module is failed. notify biomedical engineer or Mindray service staff. LOOSE CUFF Cuff is no properly LOW Properly wrap the cuff wrapped or no cuff exists. AIR LEAK Cuff, hose or LOW Check and replace the leaking parts, if connector is required, notify biomedical engineer or damaged. Mindray service staff. AIR PRESSURE Stable pressure value LOW Check if the hoses are tangled, if ERROR is not available. e.g. failure persists, notify biomedical hoses are tangled. engineer or Mindray service staff. WEAK SIGNAL Cuff is too loose or LOW Use other method to measure blood patient pulse is too pressure. weak. RANGE Measuring range HIGH Reset NIBP module, if failure persists, EXCEEDED exceeds the specified stop using NIBP measurement upper limit. function, notify biomedical engineer or Mindray service staff. EXCESSIVE Affected by arm LOW Make sure that the patient under MOTION motion, signal noise is monitoring is motionless. too large or pulse rate is not regular. OVER PRESSURE Pressure has HIGH Measure again, if failure persists, stop exceeded the using NIBP measurement function and specified upper safety notify biomedical engineer or Mindray limit. service staff. SIGNAL Excessive motion LOW Stop the patient from moving. STURATED PNEUMATIC LEAK During pneumatic LOW Check and replace the leaking parts, if test, leak is detected. required, notify biomedical engineer or Mindray service staff. NIBP SYSTEM Operation of blood HIGH Stop using NIBP measurement FAILURE pressure pump function, notify biomedical engineer or system is failed. 13-131 Mindray service staff.