Chapter 13 NIBP Monitoring 13.1 Introduction The Non-invasive Blood Pressure (NIBP) module measures the blood pressure using the oscillometric method. It is applicable for adult, pediatric, and neonatal. There are three measurement modes: MANUAL, AUTO, and CONTINUOUS. The system displays systolic, mean and diastolic pressure for each mode. In MANUAL mode, the system executes one NIBP measurement each time. In AUTO mode, the system executes NIBP measurement repeatedly with the interval of 1/2/3/4/5/10/15/30/60/90/120/180/240/480 minutes. In CONTINUOUS mode, the system executes NIBP measurement continuously within five minutes. Warning A Do not perform NIBP measurement on a patient of sickle-cell disease or with damaged skin or expected damaged skin. For a thrombasthemia patient, you must determine if to run AUTO NIBP measurement according to the clinical evaluation because haematoma may occur at the friction position between extremity and cuff. You must select correct mode especially for pediatric and neonate patients (refer to the PATIENT SETUP menu). Measurement under wrong mode may endanger patient because high adult blood pressure is not suitable for pediatric and neonate. 13.2 NIBP Monitoring 13.2.1 NIBP Measuring ZiA ^Warning /iA Use accessories specified by Mindray only, otherwise; the device may not function normally. 13-121