SpO2 Monitoring
SpO2 / Pulse Monitoring
A Warning A
ES (Electrosurgery) equipment wire and SpO2 cable must not be tangled up.
A Warning A
Do not put the sensor on extremities with arterial catheter or venous syringe.
A A Note
Do not perform SpO2 measuring and NIBP measuring on same arm at one time,
because obstruction of blood flow during NIBP measuring may adversely affect the
reading of SpO2 value.
12.2 Precautions during SpO2/Pulse Monitoring
Note
Make sure the nail covers the light window;
The wire should be on the backside of the hand.
ZiA NoteZiA
SpO2 value is always displayed at the same position.
Pulse Rate will be displayed only under following situations:
Select HR FROM as "SPO2" or "BOTH" in the ECG SETUP menu.
Select HR FROM as "AUTO" in the ECG SETUP menu and there is no ECG
signal.
Z!a Note ^
SpO2 waveform is not proportional to the pulse volume.
A A Warning
Check if the sensor cable is in normal condition before monitoring. After unplugging
the SpO2 sensor cable from the socket, the system shall display the error message
"SPO2 SENSOR OFF" and give the audible alarm.
12-112