5.5. Discontinuing a measurement
If it is necessary to interrupt a blood pressure measurement for any
reason (e.g. the patient feels unwell), the START button can be
pressed at any time.
The device immediately lowers the cuff pressure automatically.
5.6. Memory - recall of measurements
This blood pressure monitor automatically stores the last
30 measurement values. By pressing the «M» button, the
last Measurement, as well as previous measurements can M
be displayed one after the other.
mmHg mmHg mmHg mmHg
mmHg mmHg mmHg mmHg
MR MR MR MR
(MR30: Value of the last measurement) (MR29: Value of the measurement before MR 30)
5.7. Memory– cancellation of all measurements
Attention!
Before you delete all readings stored in the memory, make
sure you will not need refer to the readings at a later date.
Keeping a written record is prudent and may provide M
additional information for your doctor’s visit.
In order to delete all stored readings, depress the «M»
button for at least 7 seconds, the display will show the
symbol «CL» and 3 short beep sounds will be heard to
indicate deletion of stored readings.
6. Error Messages/Malfunctions
If an error occurs during a measurement, the measurement is
discontinued and a corresponding error code is displayed.
(Example Error no. 1)
Error No. Possible cause(s)
ERR 1 The systolic pressure was determined but afterwards the cuff pressure fell below 20 mmHg.
The tube may have become unplugged after the systolic blood pressure was measured.
Further possible cause: No pulse could be detected
ERR 2 Unnatural pressure impulses impair the measurement result. Possible cause: The arm was
moved during the measurement (artefact).
ERR 3 Inflation of the cuff takes too long. The cuff is not correctly seated or the hose connection is
not sealed.
8