Kullanım Kılavuzları
5.5. Discontinuing a measurement If it is necessary to interrupt a blood pressure measurement for any reason (e.g. the patient feels unwell), the START button can be pressed at any time. The device immediately lowers the cuff pressure automatically. 5.6. Memory - recall of measurements This blood pressure monitor automatically stores the last 30 measurement values. By pressing the «M» button, the last Measurement, as well as previous measurements can M be displayed one after the other. mmHg mmHg mmHg mmHg mmHg mmHg mmHg mmHg MR MR MR MR (MR30: Value of the last measurement) (MR29: Value of the measurement before MR 30) 5.7. Memory– cancellation of all measurements Attention! Before you delete all readings stored in the memory, make sure you will not need refer to the readings at a later date. Keeping a written record is prudent and may provide M additional information for your doctor’s visit. In order to delete all stored readings, depress the «M» button for at least 7 seconds, the display will show the symbol «CL» and 3 short beep sounds will be heard to indicate deletion of stored readings. 6. Error Messages/Malfunctions If an error occurs during a measurement, the measurement is discontinued and a corresponding error code is displayed. (Example Error no. 1) Error No. Possible cause(s) ERR 1 The systolic pressure was determined but afterwards the cuff pressure fell below 20 mmHg. The tube may have become unplugged after the systolic blood pressure was measured. Further possible cause: No pulse could be detected ERR 2 Unnatural pressure impulses impair the measurement result. Possible cause: The arm was moved during the measurement (artefact). ERR 3 Inflation of the cuff takes too long. The cuff is not correctly seated or the hose connection is not sealed. 8