M_3AS1_2_E&TR_rev2.qxd 09.04.2007 15:54 Page 8 Inflate the cuff to a pressure at least 40 mmHg higher 5.6.Discontinuing a measurement than the expected systolic pressure. If you do not know If it is necessary to interrupt a blood pressure measurement the value of this pressure, inflate to a value of for any reason (e.g. the patient feels approximately 160–180 mmHg. The cuff pressure is unwell), press the quick release valve at the backside of the shown continuously on the LCD. divece. The device then immediately lowers the cuff- pressure automatically. c) After the inflation process remain in a quiet and relaxed 5.7.Memory – displaying the last measurement sitting position. The blood-pressure monitor automatically stores the last The measurement now proceeds on its own. If no measurement value. By pressing the MEMORY button for measurement takes place and all that appears in the at least 3 seconds, the last measurement will be displayed. display is a flashing, upwards pointing arrow, see below under «Inflate further». When the instrument detects a pulse, a heart symbol in the display starts to flash and a beep is heard for every heart beat. t Further information Measurements should not occur soon after each other, since otherwise, the results will be falsified. Wait d) When the measurement has been concluded, a long therefore for at least 1 minute in a relaxed position, sitting or lying, before you repeat a measurement. peep tone sounds. The measured systolic and diastolic blood-pressure values as well as the pulse frequency are now displayed. 6. Error messages/malfunctions The cuff pressure can now be quickly released by pressing the release valve at the backside of the If an error occurs during a measurement, the measurement device. is discontinued and a corresponding error code is displayed. (example: error no. 1) 5.5.Inflating further If the cuff is not inflated sufficiently, the measurement is broken off after a few seconds and a flashing arrow pointing upwards appears in the display. It is then necessary to inflate further to at least 20 mmHg Error No. Possible cause(s) higher than the previous value. ERR 1 The systolic pressure was determined but afterwards the cuff pressure fell below 20 The instruction to inflate further can appear several times if mmHg. This situation can occur, for example, if the tube becomes unplugged after the the pressure is still not sufficient. systolic blood pressure has been measured. Further possible cause: No pulse has been detected Attention: ERR 2 The pressure dropped below 20 mmHg without a measurement taking place. The If the cuff is inflated too far (over 300 mmHg), «HI» appears Release valve was pressed during the measurement. in the display and a warning signal is sounded. The cuff ERR 5 The measured readings indicated an unacceptable difference between systolic and pressure must then be immediately released using the diastolic pressures. Take another reading following directions carefully. Contact your quick-release valve) at the pump ball)! doctor if you continue to get unusual readings. As soon as the cuff is inflated sufficiently and the measurement is proceeding, continue as described under points d–e. 8 9