Kullanım Kılavuzları
Error message Err MGM C 10 Trouble-shooting Situation NO lens filter is attached. The thermometer cannot identify a secure position of the probe. An accurate measurement was not possible. POS = position error Ambient temperature is not within the allowed operating range (10-40 50- 104 OF). Temperature taken is not within typical human temperature range (34 - 42.2 oc or 93.2-108 OF). HI = too high LO — too low System error — self-check display flashes continuously and will not be followed by the ready beep and the ready symbol. If error persists, If error still persists, Solution Attach new, clean lens filter. IRT 4020: Clear the display by pushing the "start" button once. IRT 4520: Clear the display by pushing the «I/O» button once. Take care that the positioning of the probe is correct and remains stable. Allow the thermometer to remain in a room for 30 minutes where the temperature is between 10 and 40 oc or 50 and 104 OF. Make Sure a new, clean lens filter is attached and thermometer is properly inserted. Then, take a new temperature. Wait 1 minute until the thermo- meter turns Off automatically, then turn on again. Reset the thermometer by re-moving the batteries and putting them back in. Please contact your authorized Service Centre (See Guarantee Card). Trouble-shooting Situation Battery is low, but thermometer will still operate correctly. Battery is too low to take correct temperature readings. Do you have any further questions? Replacing the batteries Solution Insert new batteries. Insert new batteries. Please contact your authorized Service Centre (See Guarantee Card). The thermometer is supplied with two 1.5 V type AA (LR 06). For best performance, we recommend Duracell@ alkaline batteries. Insert new batteries when the battery symbol appears on the display. Open the battery compartment. Remove the batteries and replace with new batteries, making sure the poles are in the right direction. Slide battery door back until it snaps in place. TO protect the environment, dispose Of empty batteries at your retail store or at appropriate collection sites according to national or local regulations. Calibration The thermometer is calibrated at the time of manufacture. If at any time you question the accuracy of the temperature measurement, please contact your authorized Service Center. Error message 11